Greisinger GPHU 014 MP/BNC pH Measurement Converter Instruction Manual

Model: GPHU 014 MP/BNC

1. Муқаддима

This manual provides comprehensive instructions for the safe and efficient operation of your Greisinger GPHU 014 MP/BNC pH Measurement Converter. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent damage. Keep this manual for future reference.

The GPHU 014 MP/BNC is designed to convert pH electrode signals into a standardized output, allowing for precise pH monitoring in various applications. It is intended for use with compatible pH electrodes featuring a BNC connector.

2. Маълумот оид ба бехатарӣ

  • Always operate the device within its specified environmental conditions.
  • Дастгоҳро ба ҳарорати аз ҳад зиёд, нурҳои бевоситаи офтоб ё намии баланд нагузоред.
  • Аз партофтан ё дучори таъсири сахти дастгоҳ худдорӣ намоед.
  • Ensure proper electrical connections are made according to the instructions to prevent electrical shock or damage.
  • Дастгоҳро худатон кушоед ё таъмир накунед. Ҳама хидматрасониро ба кормандони соҳибихтисос муроҷиат кунед.
  • Use only compatible pH electrodes and accessories with this converter.
  • Dispose of the device and any associated components in accordance with local environmental regulations.

3. Мундариҷаи баста

Лутфан, ҳангоми гирифтани баста мундариҷаи онро тафтиш кунед. Агар ягон ашё гум шуда ё вайрон шуда бошад, фавран бо таъминкунандаи худ тамос гиред.

  • Greisinger GPHU 014 MP/BNC pH Measurement Converter
  • Дастури таълимӣ (ин ҳуҷҷат)
  • Mounting accessories (if applicable, check packaging)

Note: A pH electrode is not included with this converter and must be purchased separately.

4. Маҳсулот ба охир расидview

The Greisinger GPHU 014 MP/BNC is a compact and robust pH measurement converter designed for industrial and laboratory applications. It features a clear digital display and intuitive controls for easy operation.

Greisinger GPHU 014 MP/BNC pH Measurement Converter front view

Расми 1: Пеш view of the Greisinger GPHU 014 MP/BNC pH Measurement Converter. The display shows "7.46 pH" and includes "ATC" and "Temp" indicators. Below the display are three buttons labeled "SET 1", "▼ 2", and "▲ 3". A BNC connector is visible on the right side for electrode connection. Two small ports are visible at the bottom.

Greisinger GPHU 014 MP/BNC pH Measurement Converter side view

Расми 2: Тараф view of the Greisinger GPHU 014 MP/BNC pH Measurement Converter, highlighting the BNC connector and the robust housing. The connector is labeled "AMPHENOL".

Компонентҳои асосӣ:

  • Намоиши рақамӣ: Shows pH value, temperature, and status indicators.
  • SET Button (1): Used to enter and confirm settings.
  • Тугмаи поён (2): Used to navigate menus and decrease values.
  • Тугмаи боло (3): Used to navigate menus and increase values.
  • Пайвасткунаки BNC: For connecting a compatible pH electrode.
  • Вуруди барқ: (Location not visible in images, refer to device for exact position)

5. Танзимот

5.1. Монтаж

The converter can be mounted using the integrated mounting points (screws visible in Figure 1). Ensure the mounting location is stable, free from excessive vibration, and within the specified operating temperature range.

5.2. Connecting the pH Electrode

  1. Ensure the converter is powered off.
  2. Carefully connect your pH electrode's BNC connector to the BNC input on the right side of the GPHU 014 MP/BNC converter (refer to Figure 1 and 2).
  3. Ensure the connection is secure but do not overtighten.

5.3. Пайвастшавии барқ

Connect the appropriate power supply to the converter. Refer to the device labeling for specific voltage and current requirements. Ensure the power source is stable and matches the device's specifications.

5.4. Калибрченкунии ибтидоӣ

Before first use, or after prolonged storage, the pH converter and electrode system must be calibrated. Refer to your pH electrode's manual for specific calibration buffer requirements.

  1. Қувваи барқро дар дастгоҳ.
  2. Access the calibration menu using the SET button. (Specific steps may vary, consult the device's on-screen prompts or a more detailed technical manual if available).
  3. Follow the on-screen instructions to perform a 2-point or 3-point calibration using standard pH buffer solutions (e.g., pH 4.01, 7.00, 10.01).
  4. Rinse the electrode thoroughly with distilled water between buffer solutions.
  5. Confirm calibration settings.

6. Амалиёт

6.1. Гирифтани андозагирӣ

  1. Ensure the pH electrode is properly connected and calibrated.
  2. Immerse the pH electrode into the sample solution, ensuring the sensing bulb and reference junction are fully submerged.
  3. Allow sufficient time for the reading to stabilize on the digital display. The display will show the current pH value.
  4. Note any temperature readings if the electrode has an integrated temperature sensor and ATC (Automatic Temperature Compensation) is active.

6.2. Навигатсия меню

The SET, ▼, and ▲ buttons are used to navigate through the device's menu options. Press SET to enter the menu or confirm a selection. Use ▼ and ▲ to scroll through options or adjust values.

  • SET (1): Enter/Exit menu, confirm selection.
  • ▼ (2): Scroll down, decrease value.
  • ▲ (3): Scroll up, increase value.

Specific menu functions (e.g., setting alarm limits, output configuration) will be detailed in a more advanced technical manual or on-screen prompts.

7. Нигоҳдорӣ

7.1. Cleaning the Converter

Wipe the exterior of the converter with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure no liquid enters the device.

7.2. Нигоҳдории электроди pH

Proper maintenance of the pH electrode is crucial for accurate readings and extended lifespan. Refer to your specific pH electrode's instruction manual for detailed care instructions. General guidelines include:

  • Always keep the electrode's sensing bulb hydrated, typically in a storage solution or pH 7 buffer. Never store dry.
  • Rinse the electrode with distilled or deionized water after each measurement.
  • Clean the electrode regularly to remove sample residues. Specific cleaning solutions may be required depending on the application.
  • Recalibrate the electrode regularly, especially if readings become unstable or inaccurate.

7.3. Нигоҳдорӣ

When not in use for extended periods, store the converter in a clean, dry environment at room temperature. Ensure the pH electrode is properly stored according to its manufacturer's recommendations.

8. Баррасии мушкилот

Мушкилот Сабаби эҳтимолӣ Ҳалли
Дисплей нест/Дастгоҳ фаъол нест No power, faulty power supply, incorrect connection. Check power cable and source. Ensure correct voltage.
Unstable or inaccurate pH readings Electrode not calibrated, dirty electrode, damaged electrode, air bubbles, incorrect temperature compensation. Recalibrate electrode. Clean electrode. Replace electrode if damaged. Ensure electrode is fully submerged. Check ATC settings.
Паёми "Хато" дар экран Electrode fault, out of range measurement, internal device error. Check electrode connection. Try another electrode if available. Refer to specific error codes in a detailed technical manual. Contact support if persistent.
Вақти вокуниши суст Aged or dirty electrode, low sampҳарорат. Clean or recondition electrode. Allow electrode to equilibrate with sample temperature. Consider replacing electrode.

If the problem persists after attempting these solutions, please contact Greisinger customer support or your local distributor.

9. Мушаххасоти

Хусусият Тафсилот
Истеҳсолкунанда ГРЕЙСИНГЕР
Рақами модели ашё 602331
Андозаҳои маҳсулот (L x W x H) 5.5 x 8.2 x 8 см; 9.07 грамм
ASIN B00ICKLET6
Миқдори ашё 1
Battery(ies) included Не
Battery(ies) required Не
Санаи дастрасии аввал дар Amazon.com.be 12 августи 2022
Discontinued by manufacturer Не
Мавҷудияти қисмҳои эҳтиётӣ Маълумот дар бораи қисмҳои эҳтиётӣ мавҷуд нест

10. Кафолат ва дастгирӣ

Greisinger products are manufactured to high-quality standards. For information regarding warranty terms and conditions, please refer to the warranty card included with your product or visit the official Greisinger webсайт.

For technical support, troubleshooting assistance beyond this manual, or to inquire about spare parts and service, please contact your authorized Greisinger dealer or the manufacturer directly. Contact information can typically be found on the product packaging or the official Greisinger webсайт.

Ҳуҷҷатҳои марбут - GPHU 014 MP/BNC

Пешview Дастури истифодабарии GREISINGER GE xxx: Электродҳои pH
Дастури мукаммали амалиётӣ барои электродҳои pH силсилаи GREISINGER GE xxx, ки бехатарӣ, асосҳо, андозагирӣ, калибрченкунӣ, нигоҳдорӣ, интихоб ва мушаххасотро дар бар мегирад. Роҳнамои муфассалро барои ченкунии дақиқи рН дар бар мегирад.
Пешview GREISINGER EBUW 232 EASYBUS Protocol Converter Operating Manual
Operating manual for the GREISINGER EBUW 232 EASYBUS protocol converter, detailing its specifications, general information, and safety requirements.
Пешview GREISINGER GIA 20 EB / PK Manual: Connection and Operation
This manual provides comprehensive instructions for the GREISINGER GIA 20 EB / PK microprocessor-controlled displaying, monitoring, and controlling device. It covers safety regulations, electrical connections for various input signals and output types, device configuration using software, switchpoint and alarm settings, offset/slope adjustment, serial interface, error codes, specifications, and disposal notes.
Пешview GREISINGER GMH 3710 Precision Thermometer Operating Manual
This operating manual provides detailed instructions for the GREISINGER GMH 3710 precision thermometer. It covers intended use, safety precautions, product specifications, handling, configuration, output options, input adjustments, probe connections, basic temperature measurement principles, fault messages, and disposal guidelines for Pt100 temperature sensors.
Пешview Манометрҳо ва барометрҳои силсилаи Greisinger GDH 200: Тафсилоти техникӣview
Асбобҳои ченкунии фишори дастӣ силсилаи Greisinger GDH 200-ро кашф кунед. Дар ин ҳуҷҷат манометри дақиқи GDH 200-07, манометри GDH 200-13 ва вакуум/барометри GDH 200-14 муфассал шарҳ дода шудаанд ва хусусиятҳо ва хусусиятҳои онҳоро барои истифодаи касбӣ нишон медиҳанд.
Пешview GREISINGER GMH 38-LW1 / -LW2 Moisture Meter Operating Manual Appendix
Operating manual appendix for the GREISINGER GMH 38-LW1 / -LW2 moisture measuring set for agricultural applications. Covers intended use, safety, product description, handling, principles of measurement, and specifications for wood chips, grain, hay, and straw.